Available Probes: P1 & P2
Water Level Measurements
-
The zero measurement point on both P1 and P2 Probes is the tip.
-
Feed the cable into the well.
-
The light and buzzer activate when the probe tip enters water. To ensure accuracy, lower and raise the probe a few times and then record the depth measurement from the cable at the top of the well.
Equipment Check
-
Turn sensitivity dial fully clockwise.
Notes: |
- Clockwise rotation of sensitivity dial turns meter on and increases sensitivity.
- Always set switch to highest sensitivity position, then decrease if necessary.
|
-
Depress the Battery Test button to test the battery and circuitry (excluding the probe when not in use).
-
Submerse the probe in tap water. This completes the circuit and activates the buzzer and light.