11 Diagnostics Utility
The Levelogger Diagnostics Utility can be used for troubleshooting Leveloggers and obtaining information about that Levelogger that can assist the Solinst Technical Support representative to identify and fix any problems you may encounter with your Levelogger. The Diagnostics Utility can be launched from the Utilities menu at the top of the software. The utility can be used to run a selftest, do a Memory Dump and create and email reports.
11.1 Run Diagnostics
The Run Diagnostics function reads the following information from the Levelogger:
1) Serial Number
2) Model Number
3) Firmware Version
4) Battery Voltage
5) Charge Level
6) Current Level Reading
7) Current Temperature Reading
8) Max/Min Pressure Reading
9) Max/Min Temperature Reading
10) Max/Min Conductivity Reading
11) Total Number of Logs
This information can be used to identify firmware, battery and/or temperature/ pressure sensor problems. This function also performs a series of self-tests on the Levelogger to check for problems with the battery, memory, and pressure/ temperature sensors. If any of these tests fail then a report should be created and emailed to Solinst Technical Support. To execute this function simply click the 'Run Diagnostics' button.

Figure 11-1 Run Diagnostics
(Click image to enlarge to full size)
|
|