6 Introduction to Conductivity Calibration
Start the Calibration Wizard by clicking the Conductivity Cal tab and retrieve the settings from the attached LTC Levelogger Junior by clicking .
Stop the LTC Levelogger Junior by clicking .
The wizard will guide you through the calibration process. You can exit the wizard after any step to cancel the calibration session.
Introduction: The LTC Levelogger Junior must be connected to the computer with an Optical Reader during the entire calibration process and must be in the stopped mode. Ensure you use fresh calibration solution and allow thermal equilibration of the solution during the process. Also ensure no air bubbles are present on the sensor and that the pins are clean.