|
|

Calibration Instructions
Notes: |
-
Calibrate only with 1,413 µS/cm and 80,000 µS/cm solutions. Calibrating with other solutions will cause errors in readings.
-
Do not let the probe rest on the bottom of the calibration cylinder.
|
Regular probe calibration is necessary. If, at any time, readings appear to be less accurate than usual, clean the probe thoroughly, then re-calibrate. A two point calibration should be performed prior to each measuring session. The TLC Meter automatically determines each solution being used. When measuring water of less than 10,000 µS/cm, better accuracy can be obtained if a further single point calibration at 1413 µS/cm is performed after the usual 2 point calibration.
Caution: |
The 1413 µS/cm solution can be seriously affected by salts left on the probe. Clean probe thoroughly before 1413 µS calibration by rinsing in de-ionized water until the conductivity reading reaches about 20 µS. |
- Turn the meter on. Conductivity should be displaying EC 0000u.
- Start with the 1413 µS/cm standard solution (Solinst P/N 101582).
- Clean probe and sensors as described above and give a final rinse with the calibrating solution you are about to use.
- Replace the shroud by rotating until it seats, then push to lock in place.
- Immerse the probe in the calibration solution. Press and hold the ON button until the next screen appears as Press 2x for Off.
- Again press and hold the ON button until the next screen appears as Press 2x to Cal.
- Press the ON button 2 times quickly and Cal. Now Wait appears. The probe is now being calibrated to the calibrating solution. After a few seconds, the calibration is complete and the LCD returns to the main EC and T display.
- Repeat Steps 3-7 above, using the 80,000 µS/cm (Solinst P/N 101417) calibrating solution.
- Your Solinst Model 107 TLC Meter is now calibrated and ready to use.
Notes: |
The displayed reading should be within 2% of the calibration standard being used, if not, repeat calibration. (For 1413, ± 2% = 1384 to 1441 µS/cm, for 80,000 ± 2% = 78,400 to 81,600 µS/cm |
Notes: |
- The de-ionized water, calibration solutions and the probe should all be at room temperature when conducting the calibration.
- Always ensure that no bubbles are trapped inside the probe shroud. Air bubbles will result in inaccurate calibrations.
- A repeat of the full calibration procedure.
|
Equipment Check
Upon receipt of your Solinst Model 107 TLC Meter, and always before heading out to the field, the following checks are recommended:
- Turn the meter on. If the battery is low a LOW BATT warning appears and the 9 volt alkaline battery should be replaced.
- Ensure the probe tip and shroud are clean.
- Calibrate the probe. (See calibration instructions)
Tape Guide Instructions
- Fit the Tape Guide over the top of the well, small end in.
- Insert the leg of the TLC Meter into the hole on the Tape Guide and rest the TLC Meter on the side of the well casing. (See Installation diagram.)
- Take all measurements at the V notch on the Tape Guide, and adjust readings according to the offset stamped on the Tape Guide.
- When finished, store the Tape Guide by clipping it onto the support bracket on the back of the TLC Meter.
|