Single or Straddle Packers
Solinst packers are ideal for use with Bladder Pumps or Double Valve Pumps, which can be easily attached above the packers. The water inlet can be below a single packer, or through perforated pipe fitted between straddle packers. For hydraulic conductivity testing, Solinst Leveloggers can also be suspended, either below or between packers, from the eyebolt on the bottom of the packer.

Packer Specifications
|
Packer Size OD |
1.8" (46 mm) |
3.7" (94 mm) |
Access ID |
1/2" (12.7 mm) |
1" (25.4 mm) |
Gland Length |
23" (584 mm) |
30" (762 mm) |
Overall Length |
29" (737 mm) |
36" (914.4 mm) |
- with centralizers |
Same |
44" (1,117.6 mm) |
Borehole Size |
1.9 - 2.4" (48 - 61 mm) |
3.8-4.4" (96-112 mm) |
- with centralizers |
2.5 - 3.5" (63 - 89 mm) |
4.5-5.0" (114-127 mm) |
Pipe Fittings |
1/2" NPT Female |
1" NPT Male |
- with centralizers |
Same |
1" NPT Female |
Well Size
Inflation Pressure |
1.9" - 2.4" (48 - 61 mm)
25 - 30 psi
140 - 205 kPa |
3.8 - 4.4" (96 - 112 mm)
20 psi
140 kPa |
With Centralizers |
Well Size
Inflation Pressure |
2.5 - 3.5 (63 - 89 mm)
35 - 40 psi
240 - 275 kPa |
4.5 - 5.0 (114 - 127 mm)
25 - 30 psi
170 - 205 kPa |
Notes:
-
Inflation pressures must be added to hydrostatic pressure at the packer location.
-
Maximum submergence = 150 ft (46 m)
-
Maximum pressure for 1.8" (46 mm) packer, = 50 psi or 345 kPa
above hydrostatic pressure. Maximum pressure for 3.7" (94 mm)
packers = 30 psi or 205 kPa above hydrostatic pressure.
eg: Required pressure for: Packer placed 100 ft (30 m) below water
• 1.8" packer in 2" well:
20 psi min + (100 ft x 0.43 psi) = 20 + 43 = 63 psi.
(46 mm packer: 140 kPa min + (30 x 9.8 kPa) = 140 + 294 = 434 kPa)
-
1 ft water = 0.43 psi, 1 m water = 9.8 kPa
|
|
Accessories and Optional Equipment
-
Perforated Straddle Pipe
-
Inflation Valve Assembly
-
Inflation Tubing 1/8" OD LDPE
-
Tag Line (Model 103 - marked support cable)
-
Water Level Meter (Models 101, 102)
-
-
Leveloggers (Model 3001)
-
Inflation Pump
|