solinst  peristaltic pumps  vacuum pumping  pressure delivery of liquids  pressure delivery of gases  power  operation  controls  variable speed dial  battery  operating instructions  peristaltic pump operating instructions  410 instructions  410 peristaltic pump instructions  solinst 410  solinst peristaltic pump image
  .
solinst  peristaltic pumps  vacuum pumping  pressure delivery of liquids  pressure delivery of gases  power  operation  controls  variable speed dial  battery  operating instructions  peristaltic pump operating instructions  410 instructions  410 peristaltic pump instructions  solinst 410  solinst peristaltic pump image  

Model 410 Peristaltic Pump

solinst peristaltic pumps vacuum pumping pressure delivery of liquids pressure delivery of gases image
solinst peristaltic pumps vacuum pumping pressure delivery of liquids pressure delivery of gases image


Peristaltic Pump Operating Instructions

Model 410

IMPORTANT

  • Pump is water resistant but not waterproof. Do not submerge in water.
  • Pumping rates will decrease as the amount of lift increases.
  • Remove tubing when stored.
  • Replace the tubing regularly. The pump will be inefficient if the tubing is worn. This will show as a limited ability to lift water. If the tubing looks worn, flattened or cracked, replace the tubing.
  • If used in cold weather, ensure that there is no ice in the tubing before starting the pump.
  • Pump is not designed for long-term (more than 8 hours) continuious uninterrupted use.

Power

The pump operates from an external 12V DC power supply such as a car, truck or marine 12 volt battery and has a 12 ft (3.6 m) power cable with connector clips for direct battery connection.

The power cable clips are oversized for use with automotive batteries. The red clip connects to the positive (+) battery terminal, black to negative (-) battery terminal. If the battery is connected with reverse polarity the pump will not be harmed, but it will NOT operate until the polarity is connected correctly.

An externally accessible fuse holder is located on the side of the pump case. In the event of a blown fuse due to a stalled pump head, replace fuse with an 8 AMP, type 3AG (1/4” x 1 1/4”) ‘Slo-Blow’ fuse.

Do not use a larger amperage fuse.

If the pump is to be powered by a vehicle battery for more than 3 continuous hours, start vehicle and run for 15 minutes to recharge the battery.

top 


Operation

  • The Solinst Peristaltic Pump has an integral Forward/Reverse, variable flow dial with an OFF position, which has a central detent. When turning the unit OFF, ensure you feel the dial snap into position.
  • The pump should always be kept with the control in the OFF position when it is being attached to a power source or stored.
  • Once attached to a power source, turn the dial in the direction required and adjust the flow as desired.
  • The standard 5/8” (16 mm) OD medical grade silicon tubing can be attached to 1/2” (13 mm) OD down-hole sample tubing and gives purge rates up to 3.5 L/min and 120 mL/min at the low end.
  • The optional 3/8” (10 mm) OD silicon tubing and adaptor kit allow the use of 1/4” (6 mm) OD down-hole sample tubing, giving flow rates up to 900 mL/min and sampling rates as low as 40 mL/min.

  1. Ensure that the chosen silicon tubing has been properly installed in the pump head.

  1. Connect one end of the silicon tubing to the down-hole sample line.
  2. Either connect the other end of the silicon tubing to a discharge tube, or simply discharge out of this end of the tubing.
  3. Connect the negative (black) battery clip to the negative terminal post on a 12V DC battery and the positive (red) battery clip to the positive terminal.

top 

 
solinst  peristaltic pumps  vacuum pumping  pressure delivery of liquids  pressure delivery of gases  power  operation  controls  variable speed dial  battery  operating instructions  peristaltic pump operating instructions  410 instructions  410 peristaltic pump instructions  solinst 410  solinst peristaltic pump image
solinst  peristaltic pumps  vacuum pumping  pressure delivery of liquids  pressure delivery of gases  power  operation  controls  variable speed dial  battery  operating instructions  peristaltic pump operating instructions  410 instructions  410 peristaltic pump instructions  solinst 410  solinst peristaltic pump image
.